Settings Conflict;Auto Scan Time/Meas Time Do Not Apply In Stepped Scan Type - Keysight X-Series Manual

Signal analyzers, instrument messages
Hide thumbs Also See for X-Series:
Table of Contents

Advertisement

Instrument Messages
Event Messages
Err#
Message
-221
Settings conflict;Auto
Scan Time/Meas Time do not
apply in Stepped Scan
Type.
-221
Settings conflict;Cannot
accept time or count input
when step duration type is
Continuous on step #
-221
Settings
conflict;ClearList &
Start not available when
ScanSeq = (Re)measure
-221
Settings conflict;Freq >
1 GHz is not available
while RF Input 2 enabled
-221
Settings
conflict;Function not
available while
measurement is running
-221
Settings
conflict;function
unavailable with this EMC
Standard
-221
Settings conflict;Include
Source is Yes. ARB memory
operation is rejected
-221
Settings conflict;Last
screen cannot be deleted
-221
Settings conflict;Log
Percent Auto Step Rule
does not apply in Swept
Scan Type.
-221
Settings
conflict;Multi-Screen
requires >1 screen
-221
Settings conflict;QPD +
EMI Average + RMS Average
is not allowed
Instrument Messages
Verbose/Correction Information
The Auto Scan Time/Meas Time are not available when Scan Type
=Stepped Scan.
An attempt to set list sequence a step's "Duration Time" or "Play
Count" when that step's "Duration Type" is "Continuous". This
error can only occur on SCPI input.
Set that step's "Duration Type" to "Play Count" or "Duration Time"
first then setup count or time.
Cannot perform Clear List & Start during (Re)measure because
we needs the frequency information of the peaks in signal list to
perform (Re)measure.
Frequency is limited to 1GHz while RF Input 2 is enabled.
Settings change is not allowed while measurement is running.
You must stop the current measurement if you wish to change the
settings.
The you tried to load or delete files to/from ARB memory in the
Sequence Analyzer mode when Include Source is set to Yes.
First set Include Source to No, then perform desired ARB memory
operation.
There must be at least one Screen configured, you cannot delete
the last remaining screen.
The Log Percent rule is not available when Scan Type=Swept
Scan because we are always doing linear sweep.
There must be more than one Screen configured for Multi-Screen
to be enabled.
User is not allowed to turn on all 3 EMI detectors together. You
must turn off one of the EMI Detectors before you turn this on.
66