Instrument Verification; Set Ac Line Frequency And Date/Time On The B2902A; More About Ip Addresses And Dot Notation - Keysight PD1500A System Installation Manual

Double-pulse test rack and safety enclosure
Hide thumbs Also See for PD1500A:
Table of Contents

Advertisement

Step 3: Set Instrument VISA Addresses

Instrument Verification

Set AC Line Frequency and Date/Time on the B2902A

The Keysight B2902A Source Measure Unit (SMU) is used only for the system
calibration. Its factory default line frequency is set to 50 Hz and must be changed
to 60 Hz in those countries with a 60 Hz line frequency. To change the line
frequency from the instrument front panel, press the following keys:
Optional: To set the Date and Time press the following:

More about IP Addresses and Dot Notation

Dot-notation addresses ("nnn.nnn.nnn.nnn" where "nnn" is a byte value from 0 to
255) must be expressed with care, as most PC web software interprets byte
values with leading zeros as octal (base 8) numbers. For example,
"192.168.020.011" is actually equivalent to decimal "192.168.16.9" because
".020" is interpreted as "16" expressed in octal, and ".011" as "9".To avoid
confusion, use only decimal values from 0 to 255, with no leading zeros.
Keysight Double-Pulse Test System Installation Guide
– Each of the test instruments should appear in the list of instruments.
– Run Interactive IO to verify communication with the instruments.
– For setting to 50 Hz: More > System > PLC > 50 Hz
– For setting to 60 Hz: More > System > PLC > 60 Hz
More > System >More > Info > Date/Time
Turn-on and Software Installation
53

Advertisement

Table of Contents
loading

Table of Contents