Settings Conflict; Can't Auto-Couple Res Bw In Zero Span - Keysight X-Series Manual

Signal analyzers, instrument messages
Hide thumbs Also See for X-Series:
Table of Contents

Advertisement

Instrument Messages
Event Messages
Err#
Message
-221
Settings conflict; Cannot
delete current step,
minimum number of steps
reached
-221
Settings conflict; Cannot
insert more steps,
maximum number of steps
reached
-221
Settings conflict; Cannot
optimize while user cal in
progress
-221
Settings conflict; Can't
Auto-Couple Res BW in Zero
Span
-221
Settings conflict;
Carrier freq not allowed
with BMT.
(Bottom/Middle/Top only)
-221
Settings conflict; Code
channel duplication
-221
Settings conflict; Cont
Mode is not available
-221
Settings conflict;
Continuous Peak is not
available with Fixed
marker
-221
Settings conflict;
Continuous Peak is not
available with Signal
Track on
-221
Settings conflict;
Correction file not
specified
-221
Settings conflict;
Correction not available
with Corr Group on
-221
Settings conflict;
Current setting is not
available while input is
RF/Audio
Instrument Messages
Verbose/Correction Information
An attempt was made to delete the current step which is already
the only step in the list sequence.
An attempt was made to insert more steps into list sequence
which already contains the maximum of 1000 steps.
Optimize Preselector cannot be performed while a user cal is in
progress. The user cal performs an optimize preselector prior to
taking the noise source on/off level results for the cal data.
The resolution bandwidth cannot be set to auto while you are in
zero span (time domain).
The transmit band spur measurement only allows bottom (B),
middle (M), and top (T) channel frequencies for each supported
frequency band. The carrier frequency must be set to the bottom,
middle or top frequency of the current frequency band.
This error is reported when the given code channel overlaps other
code channel
Current measurement does not support Continuous Sweep, only
Single Sweep is available
The continuous peak feature cannot be used with a marker that is
fixed. By definition that marker value cannot change.
The continuous peak feature cannot be used while you are also
using the signal tracking function.
Correction file has not been specified. There is no file to be
removed.
Correction is not available with Correction Group turned on.
This control is specific to measurement that is not currently used.
You can only perform this operation when the measurement is
selected.
51