Measuring Noise

Measuring Noise-Like Signals Using Band/Interval Density Markers

Measuring Noise-Like Signals Using Band/Interval Density Markers
Band/Interval Density Markers let you measure power over a frequency span. The
markers allow you to easily and conveniently select any arbitrary portion of the
displayed signal. However, while the analyzer, when autocoupled, makes sure the
analysis is power-responding (rms voltage-responding), you must set all of the other
parameters.
Step
1 Set the analyzer to the
Spectrum Analyzer mode.
2 Preset the analyzer
3 Set the center frequency,
span, reference level and
attenuation.
4 Measure the total noise
power between the
markers.
5 Set the band span.
6 Set the resolution and
video bandwidths.
7 Enable the internal
50 MHz amplitude
reference signal of the
analyzer.
Action
Press Mode, Spectrum
Analyzer.
Press Mode Preset.
a. Press FREQ Channel, Center
Freq, 50, MHz.
b. Press SPAN X Scale, Span, 100,
kHz.
c. Press AMPTD Y Scale, Ref
Level, −20, dBm.
d. Press AMPTD Y Scale,
Attenuation, 40, dB.
Press Marker Function,
Band/Interval Density.
Press Band Adjust,
Band/Interval Span, 40, kHz.
a. Press BW, Res BW, 1, kHz.
b. Press BW, Video BW, 10, kHz.
Press Input/Output, RF
Calibrator, 50 MHz.
78
Notes
This enables the spectrum
analyzer measurements.
Common practice is to set the
resolution bandwidth from 1% to
3% of the measurement (marker)
span, 40 kHz in this example.
Adds a discrete tone to see the
effects on the reading. See
Figure
7-4.