Frequency Offset Using Power Sensor (FOPS) – Option 208
FOPS Settings
—
Num Points
sensor measurements between the Start and Stop frequencies. Enter one
setting and the other will be changed automatically. There must be at least
two data points.
—
Dwell /Point
steps to the next frequency and before the power sensor makes a
measurement.
Amplitude Markers
Amplitude markers are horizontal dotted lines that you can move to any
amplitude level on the screen. The two amplitude markers can be set
independently or used as a pair of delta markers. Amplitude markers can be
used as a visual comparison aid with the data trace.
How to make Amplitude Marker settings
— Press Marker.
— Then More.
— Then
— Then choose from the following:
— Then
— Enter a value in dBm for the marker using the numeric keypad, the
arrows, or the rotary knob.
— Press
— Then choose from the following:
Trace Math
Trace Math allows you to compare data traces and perform simple math. Learn
how to use Trace Math in
Memory features work as documented EXCEPT for the following:
When Trace, then Data -> Mem is pressed, if a Src Power trace is active,
then RxPower is actually stored into memory.
To Compare Src Power (Input) and Rx Power (Output):
1. Meas Rx Power
2. Store to memory
Keysight N9912-90001 User's Guide
or
Freq Step Size
Specifies a settling time (in seconds) after the internal source
Ampl
Markers.
—
OFF
Amplitude markers are NOT displayed.
—
ON
Amplitude markers ARE displayed at their previous positions.
Ampl Mkr 1 or
2.
Delta Ampl
Mkr.
—
ON
AmplMkr 2 value is relative to AmplMkr 1. When AmplMkr 1 is
moved, AmplMkr 2 also moves to maintain the AmplMkr 2 value
between them.
—
OFF
The two amplitude markers move independent of one another.
"All about Trace Math" on page
Both settings control the number of power
16-19. All Data and
12- 9