TRIGger[:SEQuence]:HOLDoff <numeric_value>
This command sets the trigger holdoff in seconds.
Syntax
Parameters
Item
numeric_value
Trigger holdoff is not applicable when the power meter is set to power sweep
NOTE
mode or frequency sweep mode.
662
TRIG
:SEQ
Description/Defaul t
The trigger holdoff in seconds.
– DEF: the default value is 1 µs
– MIN: 1 µs
– MAX: 400 ms
Units are resolved to 1 ns.
:HOLD
Space
N1911A/1912A P-Series Power Meters Programming Guide
numeric_value
DEF
MIN
MAX
?
Range of Values
1 µs to 0.4 seconds
DEF
MIN
MAX