TRIGger:HOLDoff <numeric_value>
This command sets the trigger holdoff value.
Syntax
Parameter
Item
numeric_value
Example
TRIG:HOLD 0.5
Remark
If you set a trigger holdoff value which exceeds its minimum or maximum limit, the
value will be clipped to its minimum or maximum value respectively. Error
–222,"Data out of range;Value clipped to minimum (#)" or –222,"Data out of
range;Value clipped to maximum (#)" will occur.
Keysight 8990B Programming Guide
TRIG
:HOLD
Description/Defaul t
A numeric value for the trigger holdoff:
– Default value: 1 μs
– Minimum value: 1 μs
– Maximum value: 1 s
Space
numeric_value
?
This command sets the trigger holdoff to 0.5 s.
TRIGger Subsystem
Range of values
1 μs to 1 s
8