Power Sweep Mode; Figure 2-19 Trig In And Trig Out Connection Diagram Between Power Meter And Power Source - Keysight N1913 User Manual

Epm series power meters
Hide thumbs Also See for N1913:
Table of Contents

Advertisement

2
General Power Meter Functions

Power Sweep Mode

Power sweep is generally used in power level calibration setup where the
frequency is fixed (CW frequency), and the amplitude of the power source signal is
swept. This mode can be used to characterize the flatness, linearity or gain
compression of a device under test.
This feature is only available when 8480 Series, N8480 Series, E-Series E4410, or
NOTE
E-Series E9300 sensor is connected.
Procedures
1 Connect sensor to a power source.
2 Connect power meter TRIG OUT to power source TRIG IN using a BNC cable.
Figure 2-19
3 Press
80
The same connection is applied for power source TRIG OUT to power meter
TRIG IN.
TRIG
TRIG
IN
OUT
Power Source
TRIG IN and TRIG OUT connection diagram between power
meter and power source
. The Channel Setup screen displays as below.
(optional connection)
Average Power Sensor
Keysight N1913/1914A User's Guide
TRIG
TRIG
OUT
IN
Power Meter

Advertisement

Table of Contents
loading

This manual is also suitable for:

1914a

Table of Contents