Remote Interface Reference
4

Statistics operation commands (AVERage)

– You must have set CALC:FUNC AVER and CALC ON to use these commands.
CALCulate:AVERage:MINimum?
This command returns the minimum value found during a math average
operation. The meter clears the value when math is turned on, when power has
been off, or after a remote interface reset. [volatile memory]
CALCulate:AVERage:MAXimum?
This command returns the maximum value found during a math average
operation. The meter clears the value when math is turned on, when power has
been off, or after a remote interface reset. [volatile memory]
CALCulate:AVERage:AVERage?
This command returns the average of all readings taken since math average was
enabled. The meter clears the value when math is turned on, when power has
been off, or after a remote interface reset. [volatile memory]
CALCulate:AVERage:COUNt?
This command returns the number of readings taken since math average was
enabled. The meter clears the value when math is turned on, when power has
been off, or after a remote interface reset. [volatile memory]
CALCulate:AVERage:PTPeak?
This command returns the peak-to-peak value of all readings taken since math
was enabled. The meter clears the value when math is turned on, when power has
been off, or after a remote interface reset. [volatile memory]
CALulate:AVERage:SDEViation
This command returns the standard deviation of all readings taken since math was
enabled. The meter clears the value when math is turned on, when power has
been off, or after a remote interface reset. [volatile memory]
Keysight 34420A User's Guide
165