Troubleshooting

System Recovery

System Recovery
By executing system recovery, you can return the system of the E5080B (the
Windows operating system and the firmware) to the factory state (at the time
of purchase
Strictly follow the steps described below. If you do any operation other
than the following steps, the system may not be recovered.
Notes on executing factory recovery function
Executing the factory recovery function causes the following conditions:
o
o
o
Files that you have created using the save function (files in the D drive) are not
affected, but we recommend backing them up before executing system
recovery for precautionary purposes. For more information on backup, refer to
"Backing Up the Data" as described in E5080B Online Help.
24
1
).
In addition to the Windows operating system and the firmware, the
following settings of the E5080B are returned to the factory state.
• Network setting
• GPIB setting
• Printer setting
The driver for the supported printer installed after purchase is deleted.
You need to execute initial registration again.
1. The default setting of the storage is recovered during the system recovery. For the
storage that had failed earlier and had been replaced (after purchase), its default
setting may not be the same as the factory state.