Service:sensor[1]|2:Corrections:state <Boolean - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

15
SERVice Subsystem
SERVice:SENSor[1]|2:CORRections:STATe <boolean>
This command enables/disables the voltage to corrected power conversion. It
applies to E9320 Series power sensors only.
Before setting this command to OFF, you must set the INIT:CONT command to
NOTE
OFF.
After setting this command to OFF, you must only run commands relating to the
gathering of ADC values—for example, the SERV:SENS:RADC command.
Syntax
SERV
Example
SERV:SENS2:CORR:STAT ON
Reset Condition
On reset, the value is set to ON.
708
:SENS
1
2
N1911A/1912A P-Series Power Meters Programming Guide
:STAT
:CORR
This command enables the voltage to
corrected power conversion for Channel B.
Space
0|OFF
1|ON
?