SCPI Command Reference
E4980A/AL commands
[Meas Setup] - CORRECTION - SHORT A
[Meas Setup] - CORRECTION - SHORT B
[Meas Setup] - CORRECTION - LOAD A
[Meas Setup] - CORRECTION - LOAD B

:CURRent[:LEVel]

Syntax
:CURRent[:LEVel] <numeric>
:CURRent[:LEVel]?
Description
Sets the current level for the measurement signal. If the signal is set using
:VOLTage[:LEVel], the query command returns error No. 230.
Parameter
Range
Unit
Resolution
Equivalent key
[Display Format] - MEAS DISPLAY - LEVEL
[Meas Setup] - MEAS SETUP - LEVEL

:DISPlay:CCLear

Syntax
:DISPlay:CCLear
Description
Clears errors or caution messages from the display. (No Query)
Equivalent key
No equivalent key is available on the front panel.

:DISPlay:ENABle

Syntax
:DISPlay:ENABle {ON|OFF|1|0}
:DISPlay:ENABle?
Description
Enables display updates
Keysight E4980A/AL Precision LCR Meter
1. Depending on the DC Bias level.
<Numeric>
1
0 to 100m
A
1 
343