Status:questionable:enable - Keysight N774 C Series Programming Manual

Multiport power meter
Table of Contents

Advertisement

5
4
3
2
1
0
Example:
:stat:ques:cond?
Command:

:STATus:QUEStionable:ENABle

Syntax:
:STATus:QUEStionable:ENABle<wsp><value>
Description:
Sets the bits in the Questionable Status Enable Summary Mask (QSESM) that enable the contents of the QSESR to affect the
Status Byte (STB).
Setting a bit in this register to 1 enables the corresponding bit in the QSESR to affect bit 3 of the Status Byte.
Parameters:
The bit value for the questionable enable mask as a 16-bit unsigned integer value (0 .. +65535)
The default value is 65535.
Response:
none
Example:
:stat:ques:enab 128
Command:
:STATus:QUEStionable:ENABle?
Syntax:
:STATus:QUEStionable:ENABle?
Description:
Returns the QSESM for the event register
Parameters:
none
Response:
The bit value for the QSEM as a 16-bit unsigned integer value (0 .. +65535)
Example:
:stat:ques:enab?
Keysight N774xC Series Multiport Power Meter Programming Guide
Not used
Channel 4 Summary
Channel 3 Summary
Channel 2 Summary
Channel 1 Summary
Not used
->
+0
->
+128
Channel 5 Summary
Channel 4 Summary
Channel 3 Summary
Channel 2 Summary
Channel 1 Summary
Not used
Instrument Setup and Status
32
16
8
4
2
1
3
39