Mount Calibration - Keysight 478A Operating And Service Manual

Option h83 thermistor mount manual modifications
Hide thumbs Also See for 478A:
Table of Contents

Advertisement

General Information
Specifications

Mount Calibration

Table 2
Electrical Specifications (Mount Calibration)
Parameter
Effective Efficiency
Calibration Factor
Mount Calibration Data
The calibration factors are essentially the same (within 1%) as
the Effective Efficiency for the 478A Option H83 over the 1 MHz to 1000 MHz frequency
range. Both vary less than the measurement uncertainties that apply to any calibration
data, and typically remain between 98% and 100%. Thus, the CALIBRATION SWITCH on
the 432A Power Meter may be set to 99% for most measurements made with the 478A
Option H83 Thermistor Mounts. Below 1 MHz the calibration factor typically falls off as
indicated in
Figure
1.
Figure 1
Typical VSWR and Calibration Factor vs. Frequency
4
Specification
> 95% (typically > 98%)
> 95%
Thermistor Mount Manual Modifications
478A Option H83
Condition
1 MHz to 1 GHz
1 MHz to 1 GHz