[Sense[1]]|Sense2:Detector:function <Character_Data - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

10
SENSe Subsystem
[SENSe[1]]|SENSe2:DETector:FUNCtion <character_data>
This command sets the measurement mode for the E9320 and N1920 sensor
either to average or peak sensor mode.
Syntax
SENS
SENS2
Parameters
Item
character_data
[a] When measurement mode is set to average:
– If TRIG:SOUR is set to INT1, INT2 or EXT, it is set automatically to IMM.
– INIT:CONT is set automatically to ON.
– SENS:AVER2:STAT is set automatically to OFF.
– CALC:FEED is set automatically to "POW:AVG" for all CALC blocks using the specified channel in their CALC:MATH:EXPR.
[b] When measurement mode is set to NORMal:
SENS:CORR:DCYC:STAT is set automatically to OFF.
414
:
1
DET
Description/Defaul t
Defines the measurement mode:
: sets the E9320 and P-Series sensor to
AVERage
average only mode.
: sets the E9320 and P-Series sensor to normal
NORMal
mode.
N1911A/1912A P-Series Power Meters Programming Guide
:
Space
FUNC
Range of Values
AVERage
NORMal
character_data
?
[a]
[b]

Advertisement

Table of Contents
loading

This manual is also suitable for:

N1912a

Table of Contents