Command Reference
E4981A SCPI Commands
[:SENSe]:DETector:DELay1
Syntax
[:SENSe]:DETector:DELay1 <numeric>
[:SENSe]:DETector:DELay1?
Description
Sets/Gets the waiting time for analog measurement for 120 Hz Frequency
measurement.
Parameters
Description
Range
Initial value
Resolution
Query response
{numeric}<newline><^END>
Related commands
[:SENSe]:DETector:DELay2
[:SENSe]:DETector:DELay3
Equivalent key
No equivalent key can be used on front panel.
sequence
[:SENSe]:DETector:DELay2
Syntax
[:SENSe]:DETector:DELay2 <numeric>
[:SENSe]:DETector:DELay2?
Description
Sets/Gets the waiting time for analog measurement for 1 kHz Frequency
measurement.
Parameters
Description
Range
Initial value
Resolution
Query response
{numeric}<newline><^END>
250
<numeric 1>
Waiting time for 120 Hz measurement frequency
0 to 100m
1.67m
1u
on page 250
on page 251
<numeric 1>
Waiting time for 1 kHz measurement frequency
0 to 100m
1m
1u
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter