Instrument Messages
Event Messages
Err#
Message
-221
Settings conflict; Too
many points; Stop Freq
clipped to nearest value
allowed
-221
Settings conflict; Trace
Math is not available
while Normalize is on
-221
Settings conflict;
Tracking Source
unavailable in FFT Sweeps
-221
Settings conflict;
Transition Time should be
greater than # us, Acq=#
-221
Settings conflict;
Trigger Delay should be
greater than # us, Acq=#
-221
Settings conflict;
Trigger input in use for
source synchronization
-221
Settings conflict;
Trigger is not available
with span > 0 Hz.
-221
Settings conflict;
Trigger Level is too low,
Acq=#
-221
Settings conflict; Tx
Band Spur meas does not
support this frequency
band.
-221
Settings conflict; Tx
Band Spur measurement is
not defined for mobiles.
-221
Settings conflict;
X-Scale > 255 MHz
unavailable in PvT
-221
Settings
conflict;Administrator
privileges required
Instrument Messages
Verbose/Correction Information
Range Stop Frequency has been clipped to the nearest value
The Normalize function works by doing trace manipulation, so
trace math is not available while normalization is running.
Since FFT's do not sweep, you cannot use a Tracking Source
while doing FFT's
The transition time of acquisition # should be greater than physics
required minimum time.
Increase the transition time.
The trigger delay of acquisition # should be greater than physics
required minimum time.
Increase the time of trigger delay.
If Point Trigger is being used with an external trigger input to
synchronize an external source to the analyzer, that trigger input
is unavailable for triggering.
The Trigger Level on the specified Acquisition is too low, so that
the test-set cannot be triggered correctly.
Increase the trigger level.
The transmit band spur measurement does not support all of the
commercially available frequency bands. You need to change your
selection under Mode Setup, Radio, Band to one of the supported
bands.
Only base station testing is available.
In the PvT window, X-Scale (Acq BW) cannot be greater than 255
MHz, even if Option B5X is installed.
You must be logged in with Administrator privileges to do this.
Log out and log back in as the Administrator, then restart the SA
application.