Carrier Info Table; Carrier Conf Presets - Keysight X Series Measurement Manual

Signal analyzers n9083a & w9083a multi-standard radio measurement application
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Concepts
MSR Measurement Concepts
-
Bandwidth of LTE FDD parameter set
-

Carrier Info Table

Several measurements have Carrier Info Table view in which the carrier setup
summary is displayed with measured carrier powers in the order of carrier
index in one of the view windows. Though carrier powers and some power
results such as total power or reference power are displayed in the table view,
this is basically considered as setup summary table and thus its background is
white.
One of the rows is highlighted with green. This is a selected row and it is
specified either by Select Carrier on Config Carrier menu under Freq/Chan,
Carrier Setup or by a meas local parameter, e.g. Carrier Result in ACP.
Selected row number and these parameters are not coupled. The row highlight
moves to the specified one only when either of these parameters is changed.
For example, ACP is active and the carrier table is shown with the first row
highlighted. Carrier Result is 1 now. When the meas global parameter Select
Carrier is set to 2, the highlight moves to the second row. The meas local
parameter Carrier Result remains 1 because it is not coupled with any
parameter. Then if Carrier Result is changed to 3, the third row is selected.
A vertical scroll bar appears if there are too many carriers to be displayed in the
window. The scroll bar moves so that the highlighted row is displayed in the
window when the row highlight is changed. You can control the table scroll
using those two parameters that change the row selection.

Carrier Conf Presets

The 3GPP standard for multi-standard radio requires that many tests are done
with pre-defined carrier configurations. This carrier configuration is called Test
Configuration in the 3GPP documents and there are several ones defined. For
example, a test configuration for UTRA multi carrier tests is called Test
Configuration 1a which is abbreviated as TC1a. MSR mode provides partial
presets to set up carriers in the ways specified in the standard. It also provides
presets for non standard carrier configuration which just puts single RAT
carriers as many as allowed.
Carrier Config Presets set up following parameters.
68
SEM: Band Category, Carriers, Carrier Offset, Radio Format and
SPUR: Band Category, Carriers, Radio Format
Keysight N9083A & W9083A Measurement Guide