Configuring Measurement Conditions (Display and Function Related Settings)
MEAS SETUP page

Trigger Delay Time

Functional Description
The trigger delay time feature of the E4980A/AL allows you to adjust the time
between triggering and start of measurement. When you carry out list sweep
measurement, this trigger delay time is inserted into the first place of the list.
You can set the trigger delay time within the range of 0 through 999 s
(seconds) in minimum units of 100 s.
This feature is useful when you use the E4980A/AL in conjunction with the
handler and want to trigger the E4980A/AL after stabilizing the connection of
the DUT.
To set the trigger delay time:
Step 1. Press [Meas Setup].
Step 2. Using the cursor keys, select the TRIG DLY field.
Step 3. Enter the trigger delay time using the softkeys or entry keys. If you have used
the entry keys to enter the value, the softkey labels change to unit labels (ms,
s).
Softkey
INCR ++
INCR +
DECR -
DECR --
Table 4-2
Resolution steps that apply when setting the trigger delay time with INCR ++
/ DECR --
INCR ++ / DECR -- (s)
0
1 m, 2 m, 3 m, •••, 9 m
10 m, 20 m, 30 m, •••, 90 m
100 m, 200 m, 300 m, •••, 900 m
1, 2, 3, •••, 9
Keysight E4980A/AL Precision LCR Meter
Description
Increments the trigger delay time in the resolution steps shown in
Table
4-3.
Increments the trigger delay time in the resolution steps shown in
Table
4-3.
Decrements the trigger delay time in the resolution steps shown in
Table
4-3.
Decrements the trigger delay time in the resolution steps shown in
Table
4-3.