Instrument Setup and Status
Status Reporting – The STATus Subsystem
Command:

:STATus:OPERation[:EVENt]?

Syntax:
:STATus:OPERation[:EVENt]?
Description:
Returns the Operational Status Event Summary Register (OSESR).
Parameters:
none
Response:
The sum of the results for the Channels (a 16-bit unsigned integer value, where 0
Bits
9-15
8
7
6
5
4
3
2
1
0
Example:
:stat:oper?
Command:

:STATus:OPERation:CONDition?

Syntax:
:STATus:OPERation:CONDition?
Description:
Reads the Operational Status Condition Summary Register.
Parameters:
none
34
The Status subsystem allows you to return and set details from the Status
Model.
Mnemonics
N7744C
Not used
Not used
Not used
Not used
Not used
Channel 4 Summary
Channel 3 Summary
Channel 2 Summary
Channel 1 Summary
Not used
->
+0
Decimal Value
N7745C
Not used
Channel 8 Summary
256
Channel 7 Summary
128
Channel 6 Summary
64
Channel 5 Summary
32
Channel 4 Summary
16
Channel 3 Summary
8
Channel 2 Summary
4
Channel 1 Summary
2
Not used
1
Keysight N774xC Series Multiport Power Meter Programming Guide