Interpolation; Cal On ? - Questionable Accuracy; Compatible Mode Calibrations; Save The Calibration - Keysight FieldFox A Series User Manual

Table of Contents

Advertisement

Calibration for NA, CAT, and VVM Modes

Interpolation *

Highest measurement accuracy is achieved when the frequency range or
resolution settings remain the same during the measurement as when the
FieldFox was calibrated. If these settings change after performing a calibration,
the FieldFox will interpolate the calibration so that VERY accurate
measurements continue to be made.
Interpolated Calibrations are only slightly less accurate than a calibration
performed at the measurement settings. Learn more about the relative
accuracy of FieldFox calibrations in
page
When a calibration that you performed is being interpolated, an asterisk is
added to the Cal annotation. For example: Cal ON U* is shown on the screen
when the current Response or Mechanical cal is being interpolated. An * is
never added to a CalRdy.
Cal ON ? – Questionable Accuracy
When the Output Power, Interference Rejection, or IF BW (NA Mode ONLY)
setting is changed AFTER performing a calibration, a question mark is added to
the Cal annotation. For example: Cal ON Q? is shown on the screen when the
current QuickCal is being interpolated. A ? is never added to CalRdy.
QuickCal is only available on N991xA/2xA/3xA models.
The resulting measurement accuracy depends on how much the setting has
changed. For highest accuracy, recalibrate using the new settings.

Compatible Mode Calibrations

The FieldFox can have only ONE calibration present for all modes. Because NA,
CAT, and VVM modes are very similar, a calibration that is performed in one
mode can also be applied in the other modes with the same type of
measurements (1-port or 2-port).
To apply a Cal that was performed in a different mode, press Cal 5 then select
User Cal

Save the Calibration

After performing any type of calibration, you can save the FieldFox settings
along with the calibration into a STATE (*.sta) file. These settings and
calibration can then be recalled as necessary. To learn how, see
Recalling Files" on page
140
140.
ON.
544.
"Cal ON ? – Questionable Accuracy" on
Keysight N9927-90001 User's Guide
"Saving and

Advertisement

Table of Contents

Troubleshooting

loading