Figure 1-6 Rise And Fall Time Verification Setup - Keysight 8990B Service Manual

Peak power analyzer
Hide thumbs Also See for 8990B:
Table of Contents

Advertisement

Figure 1-6
5 Generate the same pulse signal as shown in
6 Press Auto Scale on the 8990B to scale the pulse signal to the optimum
7 Turn off the video bandwidth on the 8990B.
8 Measure and record the rise/fall time of the pulse signal from the 8990B into
9 Repeat
10 Compare the recorded readings to the specifications as listed in
Keysight 8990B Service Guide
Rise and fall time verification setup
display.
Table
1-2.
step 3
to
step 8
different video bandwidth settings. Record all the readings into
the test fails, refer to
Performance Verification and Adjustments
E8257D
N1923/4A
for the pulse signal with different pulse periods and
"Adjustments"
on page 51.
8990B
step
3.
Table
Table
1
1-2.
1-2. If
33