1 mW Power Reference Level Test

Description

The 1 mW power reference is used for the calibration of 8480 Series, N8480 Series
and E-Series power sensors, and is traceable to national standards. This test uses
an N8482A power sensor to transfer the power measured on an accurately
calibrated E4419B or E4417A power meter to the DUT reference.

Equipment

– Required test equipment:
– Either of these E4419B or E4417A power meters can be used. This specific
Figure 2-1
For rear panel options, the connections will differ from the illustration shown
NOTE
here. Refer to the connector identification markings on the rear panel for further
details.
Keysight N1913B/N1914B EPM Series Power Meters Service Guide
– 1 unit of E4419B or E4417A dual channel power meter
– 1 unit of N8482A power sensor
– 11730A power sensor cable
power meter model must be used.
E4419B/E4417A
POWER METER
Ch A
1 mW power reference level test setup connection diagram