Instrument Messages
Condition Messages
Err#
Bit in
Message
status
register
159
13
Settings Alert;LO may overload IF
159
13
Settings Alert; Diff probe
mismatch; <I|Q|I,Q>
159
13
Settings Alert; Acquisition
truncated
159
13
Settings Alert; Analog Out
settings conflict
159
13
Settings Alert;I/Q
mismatch:<Differential|Input
Z|Attenuation>
159
13
Settings Alert;Parm/data
mismatch
159
13
Settings Alert;Src pwr ramp>ALC
range
Instrument Messages
Error or
More Information
Warning
W
If the sweep type is Swept, the start
frequency of the instrument is less
than 10 MHz, and you put Swept IF
Gain in Manual High, then a warning
condition is generated and remains in
effect as long as this condition exists.
In some older analyzers this was error
1109.
W
The attenuation values of the two
probes on the I and/or Q channels
differ by too much for a valid
differential reading.
W
In the Analog Demod mode, certain
extreme settings combinations will
result in a required acquisition length
in excess of the capacity of the
analyzer. Increase the AF Spectrum
RBW or the RF Spectrum RBW,
decrease the Channel BW, and/or
decrease the Demod Waveform
Sweep Time.
W
The user has manually set the Analog
Output under the Input/Output menu
to a setting that conflicts with the
current measurement. There will be
no output on the Analog Out port until
this conflict is resolved. In most cases,
simply set Analog Out to Auto for the
optimal setting.
W
The impedance, differential, or
attenuation settings for the I and Q
channels do not match. For valid I+jQ
measurements the impedance and
differential settings should be the
same on both channels and the
attenuation should match within 1 dB
W
For Bluetooth, the detected
parameters did not match the data
W
You have chosen a Power Sweep
range that exceeds the ability of the
external Source to follow without
changing mechanical attenuation.
Lower your Power Sweep range.
96