Determining Reference Values; Calibration During The Application; Automatic Self-Calibration - Scanlab RTC6 PCIe Board Installation And Operation Manual

Real time control of scan systems and lasers rtc6 software package v1.16.0
Table of Contents

Advertisement

8.10.3 Determining Reference Values

The reference values are determined by
(Command
= 0). This starts the measurement routine
for determining the current Home-In positions. These
are then the reference values for later calibrations
Command
with auto_cal(
= 1 or 3 ). Immediately after
determination, they can be read out by
and are available for customer-specific calibration,
see
Section "Customer-Specific Calibration",
page
278. The reference values are stored in the
Memory
of the RTC6 board. This guarantees that the
reference values are available even after a restart.
Notes
• After
auto_cal(Command
galvanometer scanners are in the same real
Image Field
position as before the command.
• The reference values should be determined when
the machine is set up, for example, when the
scan system is installed or exchanged. When
exchanging the scan system, for which reference
values have already been determined earlier and
read out by get_hi_pos, these reference values
can be transferred directly to the RTC6 board by
write_hi_pos.
• Reference values should be determined under
conditions (ambient temperature, load) typical
for the application and after the overall system
has fully attained its operating temperature. The
reference values should always be determined
only after a warm-up time of more than 20
minutes and not before the TempOK signal has
been activated.
• Execution of auto_cal(
lasts up to 10 seconds.
• Storing the reference values takes several ms and
interrupts the 10 µs clock cycle of the DSP. For
scan systems with clock cycle monitoring
(interlock), this should be temporarily
deactivated.
RTC6 boards
Doc. Rev. 1.0.21 en-US
8 Advanced Functions for Scan Head Control and Laser Control
auto_cal
get_hi_pos
Flash
= 0), the
Command
= 0 ) typically
8.10.4 Calibration During the
Application

Automatic Self-Calibration

Automatic self-calibration of the scan system during
an application can be executed with
Command
auto_cal(
= 1 ).
This starts a measurement routine for determining
the current Home-In positions. From the deviations
from the stored reference values (see
"Determining Reference Values", page
determined in the process, gain values and offset
values are calculated and set separately for the x axis
and y axis.
This drift compensation is effective immediately. It is
valid until:
Command
• auto_cal(
= 1 ) is called again
• it is switched off
Command
– by auto_cal(
load_program_file
– after
Notes
Command
• After auto_cal(
galvanometer scanners are in the same position
in the real
Image Field
• The calibration routine started with
auto_cal(
Command
= 1 ) typically lasts
1...2 seconds (depending on the strength of the
drift). An ASC hardware check is automatically
performed, if auto_cal(
not been executed prior to the first time call of
auto_cal(
Command
= 1 ). This extends the
execution of
auto_cal
Chapter 8.10.3
277)
= 2 )
= 1 ) the
as before the command.
Command
= 0 or 4 ) has
by a few seconds.
277

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the RTC6 PCIe Board and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Related Products for Scanlab RTC6 PCIe Board

This manual is also suitable for:

Rtc6 ethernet board

Table of Contents

Save PDF