*ESE <NRf>
The *ESE (Event Status Enable) <NRf> command sets the Standard Event
Status Enable Register. This register contains a mask value for the bits to
be enabled in the Standard Event Status Register. A 1 in the Enable
Register enables the corresponding bit in the Status Register, a 0 disables
the bit. The parameter value, when rounded to an integer and expressed
in base 2, represents the bit values of the Standard Event Status Enable
Register.

Table 17-90*ESE Mapping

Syntax
742
Table 17- 90
Bit
Weight
Meaning
0
1
Operation Complete
1
2
Request Control (not used)
2
4
Query Error
3
8
Device Dependent Error
4
16
Execution Error
5
32
Command Error
6
64
Not used
7
128
Power On
Space
*ESE
?
shows the contents of this register.