Power Meter Remote Operation
3 TRIG1:DEL:AUTO OFF
4 INIT1:IMM
5 FETC1? DEF,3
Example 5 - Making a Difference Measurement
The following program segment can be carried out on the N1912A. It
queries the lower window to make a difference measurement of
Channel A - Channel B. The expected power level and resolution
parameters are defaulted, leaving them at their current settings. Some fine
tuning of the measurement is carried out by setting the averaging, and the
trigger delay to off.
Using READ?
ABOR1
ABOR2
CONF2:POW:AC:DIFF DEF,DEF,(@1),(@2)
SENS1:AVER:COUN 1024
SENS2:AVER:COUN 1024
TRIG1:DEL:AUTO OFF
TRIG2:DEL:AUTO OFF
READ2:POW:AC:DIFF?
READ2:POW:AC:DIFF? DEF,DEF,(@2),(@1) (A second READ? query is sent
to make a Channel B - Channel A measurement using fresh measurement
data).
Using INITiate and FETCh?
ABOR1
ABOR2
CONF2:POW:AC:DIFF DEF,DEF,(@1),(@2)
SENS1:AVER:COUN 1024
SENS2:AVER:COUN 1024
TRIG1:DEL:AUTO OFF
TRIG2:DEL:AUTO OFF
INIT1:IMM
INIT2:IMM
FETC2:POW:AC:DIFF?
24