The condition bits are set and cleared under the following conditions:

Table 1-10 Bit change conditions for Operation Status

Bit
Meaning
Number
0
CALibrating
4
MEASuring
5
Waiting for
TRIGger
10
SENSe
64
Bit Number
Decimal
Weight
4
16
5
32
6 - 9
-
10
1024
11
2048
12
4096
13 to 14
-
15
-
EVENts Causing Bit Changes
This is a summary bit for the Operation CALibrating Register.
SET: At beginning of zeroing (CALibration:ZERO:AUTO ONCE) and at the
beginning of calibration (CALibration:AUTO ONCE). Also for the compound
command/query CALibration[:ALL]?, this bit is set when sensor zeroing
begins.
CLEARED: At the end of zeroing or calibration.
This is a summary bit for the Operation MEASuring Register.
SET: When the power meter is taking a measurement.
CLEARED: When the measurement is finished.
This is a summary bit for the Operation TRIGger Register.
SET: When the power meter enters the "wait for trigger" state.
CLEARED: When the power meter enters the "idle" state.
This is a summary bit for the Operation SENSe Register.
SET: When the power meter is reading data from the power sensor's EEPROM.
CLEARED: When the power meter is not reading data from the power sensor's
EEPROM.
Definition
MEASuring Summary
Waiting for TRIGger Summary
Not used
SENSe Summary
Lower Limit Fail Summary
Upper Limit Fail Summary
Not used
Not used (always 0)
N1911A/1912A P-Series Power Meters Programming Guide