OUTPut:RECorder[1]|2:FEED <data_handle>
This command specifies which measurement is sent to the recorder output
specified by the numeric value following RECorder. RECorder1 applies to
both single and dual channel power meters. RECorder2 applies to dual
channel power meters only.
Syntax
Parameters
Example
OUTP:REC2:FEED "CALC1"
Reset Condition
On reset, data_handle is set to its previous value.
N1911A/1912A P-Series Power Meters Programming Guide
OUTP
:REC
Item
Description/Default
data_handle
The CALC block specifying the
measurement to be sent to the recorder
output.
:FEED
1
2
This command sends the CALC1
measurement to recorder output 2.
OUTput Subsystem
Space
data_handle
?
Range of Values
"CALC1" or "CALC"
"CALC2"
"CALC3"
"CALC4"
8