Agilent Technologies N1911A Programming Manual page 796

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

A
Calibration Factor Block Layout
Table A-99 Calibration Factor Block Layout: N8480 Series Sensors
N8480 Series
Sensors: Calibration
Factor Block Layout
Header:
Power, low
Power, high
Number of frequency
points
Bytes per frequency
point
Frequency LSB weight
Header Total:
Cal Factor Table:
Frequency (point '0')
1
Cal factor (low power)
1
Cal factor (high power)
These table entries are repeated as shown for each frequency point
Frequency (point 'N')
1
Cal factor (low power)
1
Cal factor (high power)
A-768
No.
Contents
Bytes
2
-
2
-
2
-
1
-
2
1000
9
4
-
2
-
2
-
4
-
2
-
2
-
Data
Data
Format
Range
7.8 (signed)
–127.9 to
+127.9
7.8 (signed)
–127.9 to
+127.9
16 bit
-
integer
-
-
-
32 bit fixed
0 to Fhpb*
(2^32)
2.14
0.25 to 3
2.14
0.25 to 3
32 bit fixed
0 to Fhpb*
(2^32)
2.14
0.25 to 3
2.14
0.25 to 3
N1911A/1912A P-Series Power Meters Programming Guide
Units
Notes
dBm
Power for low power
flatness.
dBm
Power for high power
flatness.
None
None
Number of bytes in cal
factor value at each
frequency and power
level.
Hertz
Fhbp (Freq. Hz per bit).
1 KHz per bit for the cal
factor: 1 KHz x 2^32 =
4.3E+12 = 4300 GHz
range
None
Fhpb = Freq Hz per bit
None
Power (in watts) is
divided by this value.
None
Power (in watts) is
divided by this value.
None
Fhbp = Freq Hz per bit
None
Power (in watts) is
divided by this value.
None
Power (in watts) is
divided by this value.