CALCulate[1]|2|3|4:GAIN:STATe <boolean>
This command is used on the specified window/measurement to enable
and disable the display offset set by the
CALCulate[1]|2|3|4:GAIN[:MAGNitude] command.
Syntax
Example
CALC2:GAIN:STAT 1
Reset Condition
On reset, the gain is disabled.
Query
CALCulate[1]|2|3|4:GAIN:STATe?
The query enters a 1 or 0 into the output buffer indicating the status of
the display offset.
• 1 is returned when the display offset feature is enabled
N1911A/1912A P-Series Power Meters Programming Guide
CALC
:GAIN
1
2
3
4
Space
:STAT
This command enables the display offset
for the lower window/ upper
measurement.
CALCulate Subsystem
0|OFF
1|ON
?
3
171