Status:questionable:calibration[:Summary] - Agilent Technologies N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

11
STATus Subsystem

STATus:QUEStionable:CALibration[:SUMMary]

The questionable calibration summary register set contains bits which give
an indication of the quality of the data produced by the power meter due
to its calibration status.
The following bits in these registers are used by the power meter:
These bits are set by the following:
• Error –231, "Data questionable; CH<A|B>:ZERO ERROR"
• Error –231, "Data questionable; CAL ERROR"
• Error –231, "Data questionable; CAL ERROR ChA"
• Error –231, "Data questionable; CAL ERROR ChB"
These bits are cleared when any of the three commands listed above
succeed and no errors are placed on the error queue.
Syntax
486
Bit
Decimal
Number
Weight
0
-
1
2
2
4
3-15
-
STAT
:QUES
Definition
Not used
Summary of Channel A CALibration
Summary of Channel B CALibration (N1912A only)
Not used
:CAL
:SUMM
N1911A/1912A P-Series Power Meters Programming Guide