Fetch[1]|2|3|4[:Scalar][:Power:ac]? [<Expected_Value>[,<Resolution>[,<Source List>]]] - Agilent Technologies N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

FETCh[1]|2|3|4[:SCALar][:POWer:AC]?
[<expected_value>[,<resolution>[,<source list>]]]
This command sets the specified window's measurement function to single
channel with relative mode off, recalculates the measurement and places
the result on the bus. The result is a power based measurement and is
expressed in the units defined by UNIT[1]|2|3|4:POWer.
Syntax
FETC
1
2
3
4
Parameters
Refer to
parameters in this command.
N1911A/1912A P-Series Power Meters Programming Guide
:POW
:SCAL
expected_value
Space
DEF
"Optional Parameters"
Item
Description/Default
expected_value
The expected power level parameter can be
(for the expected
set to DEF or a numeric value. If a value is
power level)
entered it should correspond to that set by
CONFigure otherwise an error occurs. The
units of measurement are dBm and W. The
default units are defined by UNIT:POWer.
:AC
?
,
resolution
DEF
on page 90 for additional details on the
MEASurement Commands
,
source list
Range of Values
sensor dependent
1
DEF
2
111