Trigger[:Sequence[1]|2]:Delay:auto <Boolean - Agilent Technologies N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

14
TRIGger Subsystem
TRIGger[:SEQuence[1]|2]:DELay:AUTO <boolean>
This command is used to determine whether or not there is a
settling- time delay before a measurement is made.
When this command is set to:
• ON, the power meter inserts a settling- time delay before taking the
• OFF, no settling- time delay is inserted and the power meter makes the
• ONCE, a settling- time delay is inserted before taking the requested
TRIGger[1]|2:DELay:AUTO is ignored if TRIGger[1]|2[:IMMediate] is
set to ON.
Syntax
TRIG
680
requested measurement and for subsequent measurements. This settling
time allows the internal digital filter to be updated with new values to
produce valid, accurate measurement results. The trigger with delay
command allows settling time for the internal amplifiers and filters. It
does not allow time for power sensor delay.
In cases of large power changes, the delay may not be sufficient for
complete settling. Accurate readings can be assured by taking two
successive measurements for comparison.
measurement immediately a trigger is received.
measurement, for one measurement only.
:SEQ
1
2
:DEL
:AUTO
N1911A/1912A P-Series Power Meters Programming Guide
Space
0|OFF
1|ON
ONCE
?