Memory:table:select <Character_Data - Agilent Technologies N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

7
MEMory Subsystem
MEMory:TABLe:SELect <character_data>
This command is used to activate either a sensor calibration table (8480
Series sensors and N8480 Series sensors with Option CFT only), or a
frequency dependent offset table. A table must be activated before any
operation can be performed on it.
Syntax
Parameters
Example
MEM:TABL:SEL "Sensor1"
Query
MEMory:TABLe:SELect?
The query returns the name of the currently selected table.
290
MEM
:TABL
Item
Description/Default
character_data
Details the new table name. A maximum of
12 characters can be used.
Space
character_data
:SEL
?
This command selects a sensor calibration
table named "Sensor1".
N1911A/1912A P-Series Power Meters Programming Guide
Range of Values
A to Z (uppercase)
a to z (lowercase)
0 - 9
_ (underscore)

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1912aN1912a

Table of Contents