PSTatistic[1]|2:CCDF:COUNt <numeric_value>
This command is used to set CCDF cummulative counts for Channel A or
Channel B. The increment step for the cummulative counts is 100 M.
This command is only applicable when P-Series sensor is present and free-run acquisition
N O T E
mode is selected. If P-Series sensor is used by the adjacent channel in a dual channel
setup, the same setting will be applied for free-run acquisition mode.
Syntax
Parameters
Example
PST1:CCDF:COUN 1.2G
332
PST
:CCDF
1
2
Item
Description/Default
numeric_value
The CCDF cummulative counts in numeric
value.
• Minimum value: 100 M
• Maximum value: 10 G
Space
:COUN
This command sets the CCDF
cummulative counts for Channel A to 1.2
G.
N1911A/1912A P-Series Power Meters Programming Guide
numeric_value
?
Range of Values
100 M to 10 G