IEEE488.2 Command Reference
*ESE <NRf>
*ESE <NRf>
The *ESE (Event Status Enable) <NRf> command sets the Standard
Event Status Enable Register. This register contains a mask value for the
bits to be enabled in the Standard Event Status Register. A 1 in the
Enable Register enables the corresponding bit in the Status Register, a 0
disables the bit. The parameter value, when rounded to an integer and
expressed in base 2, represents the bit values of the Standard Event
Status Enable Register. Table 16-3 shows the contents of this register.
Bit
0
1
2
3
4
5
6
7
Syntax
*ESE
Parameters
Type
NRf
Query
*ESE?
The query returns the current contents of the Standard Event Status
Enable Register. The format of the return is <NR1> in the range of
0 to 255.
16-10
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Table 16-3: *ESE Mapping
Weight
Meaning
1
Operation Complete
2
Request Control (not used)
4
Query Error
8
Device Dependent Error
16
Execution Error
32
Command Error
64
Not used
128
Power On
Space
NRf
?
Description/Default
A value used to set the Standard Event
Status Enable Register.