Trigger[:Sequence]:Holdoff <Numeric_Value - Agilent Technologies N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

TRIGger[:SEQuence]:HOLDoff <numeric_value>
This command sets the trigger holdoff in seconds.
Syntax
Parameters
Trigger holdoff is not applicable when the power meter is set to power sweep mode or
N O T E
frequency sweep mode.
N1911A/1912A P-Series Power Meters Programming Guide
TRIG
:SEQ
Item
Description/Default
numeric_value
The trigger holdoff in seconds.
Units are resolved to 1 ns.
:HOLD
Space
DEF: the default value is 1 µs
MIN: 1 µs
MAX: 400 ms
TRIGger Subsystem
numeric_value
DEF
MIN
MAX
?
Range of Values
1 µs to 0.4 seconds
DEF
MIN
MAX
14
667