Agilent Technologies N1911A Programming Manual page 492

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

10
SENSe Subsystem
Example
SENS:TRAC:Y:SCAL:PDIV
0.002
SENS2:TRAC:Y:SCAL:PDIV
0.05
Reset Condition
On reset, Y Max = 20 dBm; Y Scale = 7 dB.
Query
SENSe[1]|2:TRACe:Y:SCALe:PDIV?
The query returns the current scale setting of Y- axis in numerical value.
Query Example
SENS:TRAC:Y:SCAL:PDIV?
Error Messages
• If no sensor or wrong sensor is connected to the channel, error –241
• If channel is not in trigger acquisition mode, error –221 "Settings
• If limits of the values keyed in are exceeded, error –222 "Data out of
464
"Hardware missing" occurs.
conflict" occurs.
range; upper (or lower) limit exceeded; no change" occurs.
This command sets the Channel A Y Scale
value to 0.002 step.
This command sets the Channel B Y Scale
value to 0.05 step.
This command queries the Channel A
current Y- axis scale setting in numerical
value.
N1911A/1912A P-Series Power Meters Programming Guide