SERVice Subsystem
SERVice:SENSor[1]|2:PCALfactor <cal_factor_data>
This command writes calibration factor data to, or reads calibration factor
data from, the currently connected sensor. The whole calibration factor
block must be written at once as a checksum is generated. The new block
must not be larger than the existing block.
This command applies to E9320 Series sensors for peak path data only.
For E4410 Series, E9300 Series and E9320 Series sensors, average path
data, refer to
page 710.
Syntax
Parameters
Query
SERVice:SENSor[1]|2:PCALfactor?
The query returns the current peak path calibration factor block.
718
"SERVice:SENSor[1]|2:CALFactor <cal_factor_data>"
SERV
:SENS
Item
Description/Default
cal_factor_data
A binary data block. Refer to
Layout," on page A-763 for further information.
Space
1
:PCAL
2
Appendix
N1911A/1912A P-Series Power Meters Programming Guide
on
cal_factor_data
?
A, "Calibration Factor Block