LLO
The LLO (Local Lock Out) command can be used to disable the front panel
local key. With this key disabled, only the controller (or a hard reset by
the line power switch) can restore local control.
PPC
When addressed to listen, the PPC (Parallel Poll Configure) command
causes the power meter to be configured according to the parallel poll
enable secondary command which should follow this command.
PPD
Sending the PPC command followed by the PPD (Parallel Poll Disable)
command disables the power meter from responding to a parallel poll.
This is effectively a selective disable.

Table 17-94PPD Mapping

N1911A/1912A P-Series Power Meters Programming Guide
Bit
Weight
Meaning
0
1
Always 0
1
2
Always 0
2
4
Always 0
3
8
Always 0
4
16
Always 1
5
32
Always 1
6
64
Always 1
7
128
Always 0
IEEE 488.2 Command Reference
17