Table 7-21 Frequency And Calibration/Offset Factor List - Agilent Technologies N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

7
MEMory Subsystem

Table 7-21 Frequency and Calibration/Offset Factor List

For sensosr calibration tables (8480 Series sensors and N8480 Series
sensors with Option CFT only), the number of frequency points must be
one less than the number of calibration factor points. This is verified
when the sensor calibration table is selected using
SENSe:CORRection:CSET:SELect <string>.
Ensure that the frequency points you use cover the frequency range of the
signals that you want to measure. If you measure a signal with a
frequency outside the frequency range defined in the table, then the power
meter uses the highest or lowest point in the table to calculate the
calibration factor/offset.
Depending on available memory, the power meter is capable of storing 20
sensor calibration tables and 10 frequency dependent offset tables, each
containing 80 points.
280
Table
Power Sensor
0
None
1
8481A
2
8482A, 8482B, 8482H
3
8483A
4
8481D
5
8485A
6
R8486A
7
Q8486A
8
R8486D
9
8487A
1
Default is a sensor calibration table in which the reference calibration factor and calibration
factors are 100%. This sensor calibration table can be used during the performance testing of the
power meter.
N1911A/1912A P-Series Power Meters Programming Guide
Table Name
1
DEFAULT
8481A
8482A
8483A
8481D
8485A
R8486A
Q8486A
R8486D
8487A