1
Power Meter Remote Operation

Setting Measurement Limits

You can configure the power meter to detect when a measurement is
outside of a predefined upper and/or lower limit value.
Limits are window or measurement display line based and can be applied
to power, ratio or difference measurements.

Setting Limits

The power meter can be configured to verify the power being measured
against an upper and/or lower limit value. The range of values that can be
set for lower and upper limits is –150.00 dBm to +230.00 dBm. The
default upper limit is +90.00 dBm and the default lower limit is
–90.00 dBm.
A typical application for this feature is shown in
Figure 1-4 Limits Checking Application
40
Swept Source
OUT
N1911A/1912A P-Series Power Meters Programming Guide
Device
Under Test
IN
OUT
Figure 1-
4.
Power Meter
CHANNEL A
INPUT