Power Meter Remote Operation
Example 1 - The Simplest Method
The following program segments show the simplest method of querying the
upper and lower window's measurement results respectively.
Using READ?
*RST
CONF1
READ1?
*RST
CONF2
READ2?
Using INITiate and FETCh?
*RST
CONF1
INIT1?
FETC1?
For the N1911A only:
*RST
CONF2
INIT1
FETC2?
For the N1912A only:
*RST
CONF2
INIT2?
18
Reset instrument
Configure upper window - defaults to a Channel A
measurement
Take upper window (Channel A) measurement
Reset instrument
Configure lower window - defaults to a Channel A (N1911A),
Channel B (N1912A) measurement
Take lower window measurement (Channel A on N1911A, B
on N1912A)
Reset instrument
Configure upper window - defaults to a Channel A
measurement
Causes Channel A to make a measurement
Retrieves the upper window's measurement
Reset instrument
Configure lower window - N1911A defaults to Channel A
Causes Channel A to make a measurement
Retrieves the lower window's measurement
Reset instrument
Configure lower window
Causes Channel B to make a measurement
N1911A/1912A P-Series Power Meters Programming Guide