READ[1]|2|3|4[:SCALar][:POWer:AC]?
[<expected_value>[,<resolution>[,<source list>]]]
This command sets the specified window's measurement function to single
channel with relative mode off, aborts then initiates the specified channel,
calculates the measurement result and places the result on the bus. The
result is a power based measurement and is expressed in the units
defined by UNIT[1]|2|3|4:POWer.
INITiate:CONTinuous must be set to OFF, otherwise error –213, "INIT ignored"
N O T E
occurs. If TRIGger:SOURce is set to BUS, error –214, "Trigger deadlock" occurs.
Syntax
READ
1
2
3
4
Parameters
Refer to
parameters in this command.
N1911A/1912A P-Series Power Meters Programming Guide
:SCAL
:POW
expected_value
Space
DEF
"Optional Parameters"
Item
Description/Default
expected_value
The expected power level parameter can be
(for the expected
set to DEF or a numeric value. If a value is
power level)
entered it should correspond to that set by
CONFigure otherwise an error occurs.
:AC
?
,
resolution
DEF
on page 90 for additional details on the
MEASurement Commands
,
source list
Range of Values
sensor dependent
1
DEF
2
129