Sense[1]|2:Trace:y:scale:pdiv <Numeric_Value - Agilent Technologies N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

SENSe[1]|2:TRACe:Y:SCALe:PDIV <numeric_value>
This command is used to set the Y Scale value (per division) for the
selected trace.
The Y Scale value set by this SCPI command is dependant on the current
Y- axis unit, which can be set using SENS:TRAC:UNIT command.
This command is only applicable when N192x or E932x sensor is present and trigger
N O T E
acquisition mode is selected.
Syntax
SENS
1
2
Parameters
N1911A/1912A P-Series Power Meters Programming Guide
:TRAC
:Y
Item
Description/Default
numeric_value
The numeric value for Y-axis scale.
:PDIV
:SCAL
SENSe Subsystem
numeric_value
Space
?
Range of Values
0.001 dB to10 dB
1 nWatt to10 MWatt
10
463

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1912aN1912a

Table of Contents