PSTatistic Subsystem
PSTatistic[1]|2:CCDF:PROBability? <numeric_value>
This command is used to return the probability at the specified power
level.
This command is only applicable when P-Series sensors are present and the following
N O T E
conditions are met:
• Free run acquisition mode is selected
• NORMal or DOUBle measurement speed setting is chosen
Syntax
Parameters
340
PST
1
2
Item
numeric_value
:PROB
:CCDF
Description/Default
The power level at the queried
probability.
• Maximum value: 50.00 dB
• Minimum value: 0.00 dB
N1911A/1912A P-Series Power Meters Programming Guide
?
space
numeric_value
Range of Values
0.00 to 50.0