CALibration[1]|2:ZERO:NORMal:AUTO <boolean>
This command provides a quick way of zeroing the NORMAL path of an
E9320 Series sensor. The average only path is unaffected. This command
can only be used to zero an E9320 Series sensor and a P- Series sensor.
The P-Series sensor only has a NORMAL path. Hence, the reason this E9320 Series sensor
N O T E
command is allowed to function.
The command causes the power meter to perform its zeroing routine, on
the specified channel, when ONCE is selected. This adjusts the power meter
for a zero power reading with no power supplied to the power sensor.
The 0|OFF parameter is only required for the query response and is
ignored in the command. If 1|ON is selected on an E9320 Series sensor, it
causes the error –224, "Illegal parameter value" to occur.
Except when using a P- Series sensor, this command assumes that the
E9320 Series sensor is not connected to a power source.
Syntax
CAL
Example
CAL2:ZERO:NORM:AUTO ONCE
N1911A/1912A P-Series Power Meters Programming Guide
:ZERO
1
2
:NORM
:AUTO
This command causes the power meter to
perform a zeroing routine on Channel B.
CALibration Subsystem
Space
0|OFF
ONCE
?
4
215