Using READ?
ABOR1
CONF1 DEF,3
READ1?
Some fine tuning of the above program segment can be carried out for
example, by setting the trigger delay off. The following program segment
assumes that Channel A is currently being measured on the upper
window.
1 ABOR1
2 CONF1 DEF,3
3 TRIG1:DEL:AUTO OFF
4 READ1?
Using INITiate and FETCh?
The following program segment assumes that Channel A is currently being
measured on the upper window.
ABOR1
CONF1 DEF,3
INIT1
FETC1? DEF,3
Some fine tuning of the above program segment can be carried out for
example, by setting the trigger delay off.
1 ABOR1
2 CONF1 DEF,3
N1911A/1912A P-Series Power Meters Programming Guide
Aborts Channel A
Configures the upper window to make a measurement
using the current setting of the expected power and
source list and a resolution setting of 3
Takes the upper window's measurement. This is
Channel A or B measurement depending on current
window setup.
Aborts Channel A
Configures the upper window to make a
measurement using the current setting of the
expected power and source list and a resolution
setting of 3
Causes Channel A to make a measurement
Retrieves the upper window's measurement
Power Meter Remote Operation