PSTatistic:CCDF:REFerence:POWer:AVERage?

This command is used to retrieve average power data of the saved
reference trace.
This command is only applicable when P-Series sensors are present and the following
N O T E
conditions are met:
• Free-run acquisition mode is selected
• NORMal or DOUBle measurement speed setting is chosen
Syntax
Example
PST:CCDF:REF:POW:AVER?
Error Messages
• If command is executed in other window besides CCDF window, error
• If there was no previously saved trace, error –221 "Settings conflict: No
N1911A/1912A P-Series Power Meters Programming Guide
PST
:CCDF
–221 "Settings conflict: Requires CCDF window" occurs.
reference trace saved" occurs. Please check the status of saved reference
trace using command PST:CCDF:STOR:REF?.
:REF
:POW
This command returns the average power
value of the reference trace.
PSTatistic Subsystem
?
:AVER
9
327